Humidity Cycling

Purpose & Application

The ENV-03-2 humidity cycling chamber executes the full range of standard cyclic damp heat tests including IS 9000-V Section 1/2, IS 9000-VI composite, JSS 55555 Test 17 and MIL-STD-810 cycles B1, B2 and B3. Auto programming controls cycle profile and duration. Single chamber method preferred over dual-chamber for most procedures.

Available Sizes

ModelWorkspace (H x W x D)Static LoadShelves
ENV-03-2-125 Liters50 x 50 x 50 cm10 Kg1
ENV-03-2-450 Liters75 x 75 x 80 cm25 Kg2
ENV-03-2-1000 Liters100 x 100 x 100 cm50 Kg2
ENV-03-2-CBCustom BuiltSpecifySpecify

We have delivered Hot Cold & Humidity Chambers up to 10 x 4.5 x 3.5 metres.

Key Specifications

Temperature Range
-15 degC to +80 degC
Relative Humidity
14% to 98% RH
Control Accuracy
+/-0.5 degC and +/-2% RH
Programmer
Eurotherm 2604 or PLC-HMI (Watlow / Eurotherm)
Sensors
Pt-100 RTD + capacitive RH sensor
Construction
SS 304 18-gauge interior, CRCA exterior, PU paint
Power Supply
415 V, 3-phase or 230 V, 1-phase, 50 Hz
Standards
IS 9000-V/VI, JSS 55555, MIL-STD-810

Built-in Capabilities

  • Auto-controlled humidity cycle profiles
  • Eurotherm 2604 or PLC-HMI programmer
  • Single-stage refrigeration system
  • Low-pressure steam humidification
  • Multilayer observation window with illumination
  • Adjustable shelves and 50/100/150 mm port holes
  • Caster wheels on models up to 1000 litres
  • Optional data logger when Eurotherm is selected

Protection & Reliability

  • Independent over- and under-temperature protection
  • Motor winding over-temperature protection on semi-sealed compressors
  • High-discharge / low-suction pressure correction before tripping
  • Compressor trip on abnormal pressure conditions
  • Over-current relays (OLR) and MCBs as required
  • Audio-visual alarm for critical functions
  • Low water level protection in humidity generator

Configure Your Chamber

Tell us your testing requirements and we'll design a chamber tailored to your specifications. Our engineers respond within 24 hours.